Trace models · OxyScout Ultra-Trace and OxyScout Trace

Trace Oxygen Analyzers for Natural Gas

OxyScout Ultra-Trace and OxyScout Trace are optical oxygen analyzers for continuous ppm-level O₂ measurement in natural gas. Ultra-Trace covers 0–50 ppm* for low-ppm specifications, and Trace covers 0–1,000 ppm*. Neither model has an electrochemical cell to replace.

What is a trace oxygen analyzer?

A trace oxygen analyzer measures oxygen in parts per million (ppm) by volume. Ambient air is about 20.9% oxygen, or 209,000 ppm, so a small leak of air into a sample line or process stream shows up quickly at ppm levels. OxyScout trace models measure with a fully optical sensor and send the reading over 4–20 mA and RS-485 Modbus RTU at the same time.

Which trace model fits the application?

ModelMeasuring range*Detection limit*Accuracy*Typical application
OxyScout Ultra-Trace0–50 ppm O₂≤0.5 ppm O₂±2 ppm or ±5% of reading, whichever is greaterSales gas, custody transfer and vapor recovery units with a low-ppm specification
OxyScout Trace0–1,000 ppm O₂50 ppm O₂Typically ±2% of reading or ±0.01% O₂ (100 ppm), whichever is greaterGathering, vapor recovery units and process

*Based on controlled lab conditions.

Choose a range that puts your oxygen limit well inside the scale, with enough headroom above the limit to see an upset rather than a pegged signal. OxyScout Trace has a 50 ppm detection limit*, so a low-ppm specification calls for OxyScout Ultra-Trace. The pipeline oxygen limits guide matches published limits to each range.

OxyScout Ultra-Trace: 0–50 ppm

OxyScout Ultra-Trace is the low-ppm model, with a 0–50 ppm O₂ range* and a detection limit of ≤0.5 ppm O₂*.

Measuring range*0–50 ppm O₂
Detection limit*≤0.5 ppm O₂
Accuracy*±2 ppm or ±5% of reading, whichever is greater
Typical applicationSales gas, custody transfer and vapor recovery units with a low-ppm specification

*Based on controlled lab conditions.

OxyScout Trace: 0–1,000 ppm

OxyScout Trace covers 0–1,000 ppm O₂* with a 50 ppm O₂ detection limit*.

Measuring range*0–1,000 ppm O₂
Detection limit*50 ppm O₂
Accuracy*Typically ±2% of reading or ±0.01% O₂ (100 ppm), whichever is greater
Typical applicationGathering, vapor recovery units and process

*Based on controlled lab conditions.

What specifications do both trace models share?

Both trace models share the installation, wiring, firmware and communications of the full OxyScout line. The full specification table also covers options and accessories.

MeasurementThe analyzer uses a fully optical O₂ sensor, so there are no electrochemical cells to replace.
OutputsThe outputs are a 4–20 mA loop (current sink) and RS-485 Modbus RTU (2-wire half-duplex).
PowerThe analyzer runs on 9–36 VDC and draws 35 mA plus the current used by its analog outputs.
Process / ambient temperatureProcess and ambient temperature is 0 to 50 °C, and storage temperature is −20 to 70 °C.
PressureUp to 250 PSI input pressure†
Humidity / flowThe analyzer is specified for non-condensing conditions, and it has no flow restrictions.
Area classificationThe analyzer is engineered for Class 1 Division 2 (Zone 2) hazardous areas.
Size / weightThe enclosure is 9 in high, 7 in wide and 2.5 in deep, and the analyzer weighs 4 lb.
Process connectionThe process connection is 1/8 in tube compression.
Response time*Typical response time is under 10 seconds.

*Based on controlled lab conditions. Standard Sensors Corporation reserves the right to make revisions and/or changes without notice.

†Without Flowmeter installed. Higher ratings if regulator installed upstream.

How should the sample line be set up for trace oxygen?

Most high readings at ppm levels come from air entering the sample system, not from the analyzer. Oxygen moves inward through any imperfect fitting against line pressure, because each gas moves according to its own partial pressure.

  • Run the sample from tap to analyzer in stainless steel tubing only, because plastic and polymer tubing lets atmospheric oxygen diffuse through the tube wall.
  • Use metal-diaphragm regulators and metal-sealed components everywhere upstream of the analyzer.
  • Keep the run as short as practical, because a smaller trapped volume gives a faster response.
  • Leak-check every fitting from the analyzer back to the tap with a liquid leak detector while the line is pressurized.
  • Let a new installation purge until the reading has stopped falling before you calibrate, which can take several hours.

The ZeroLeak™ Cell Block on every OxyScout is machined from one solid block, so minimal fittings are required inside the analyzer. Read more about the ZeroLeak™ Cell Block.

Our team is trusted by
Energy Transfer logoHess logoOxy logoExxonMobil logoChevron logo
FAQ

Frequently asked questions

What is the lowest oxygen level OxyScout can detect?

OxyScout Ultra-Trace has a detection limit of ≤0.5 ppm O₂.* OxyScout Trace has a detection limit of 50 ppm O₂*, so Ultra-Trace is the model for a low-ppm specification. *Based on controlled lab conditions.

Why does a new trace installation read high at first?

Air trapped in tubing, fittings and regulators bleeds slowly into the flowing sample. On a fresh installation, expect the reading to fall steadily for several hours, and longer on long or complex sample runs, before it reaches a stable low-ppm value.

Does H₂S or CO₂ in the gas affect the reading?

OxyScout has no cross-sensitivity or poisoning from H₂S, CO₂ or the other gases that plague galvanic cells.